Peristaltic Pumps Singapore

Published: {{NEW_ARTICLE_DATE}}  |  By JWRESEARCH  |  Category: Laboratory Equipment

A peristaltic pump is the quiet workhorse behind a lot of laboratory work: feeding media to a bioreactor, dosing reagent into a titration, sampling from a vessel that must not be opened, or simply moving a suspension that would destroy an ordinary pump. Because the fluid only touches the inside of a flexible tube, the pump never has to be cleaned between samples — you change the tube and carry on.

At JW RESEARCH SUPPLY, we supply laboratory peristaltic pumps and tubing to research groups, contract laboratories and industrial customers across Singapore. This guide covers the pump types, how to match flow to tubing, dosing accuracy and the maintenance rules that keep a pump honest.

How a Peristaltic Pump Works

Rollers mounted on a rotating head compress the tube progressively, pushing a slug of liquid ahead of each roller while the tube behind it recovers and refills. Nothing else in the system is wetted: no valves, no seals, no impeller. That gives three practical advantages — no cross-contamination between samples, tolerance of suspensions and slurries, and the ability to run dry without damage. The cost is gentle pulsation in the flow, modest pressure capability, and tubing that wears and must be replaced on a schedule.

Pump Types Compared

Type Typical Flow Range Channels Best For
Fixed-speed transfer pump ≈ 10 – 300 mL/min 1 Simple media and reagent transfer
Variable speed pump ≈ 0.1 – 500 mL/min 1 – 2 Adjustable flow, perfusion and recirculation
Multichannel pump ≈ 0.01 – 100 mL/min per channel 4 – 12 Parallel dosing, plate work, media preparation
Programmable dosing pump ≈ 0.001 – 200 mL/min 1 – 4 Timed or volume-based dispensing, calibration routines
Micro peristaltic pump ≈ 0.001 – 10 mL/min 1 – 8 Microfluidics, droplet work, low-flow sampling

Specification Reference

Parameter Typical Value Notes
Tubing internal diameter 0.5, 0.8, 1.6, 3.2, 4.8, 6.4, 8.0 mm Flow scales with bore; 1.6 and 3.2 mm cover most bench work
Tubing wall thickness 1.6 mm or 2.4 mm Must match the pump head; mismatched wall ruins flow accuracy
Speed range ≈ 1 – 300 rpm Low-speed resolution matters more than peak speed
Max pressure ≈ 1 – 3 bar Well below a diaphragm pump; plan the flow path accordingly
Dosing accuracy ≈ ± 1 – 3 % after calibration Verify in your own tubing, liquid and speed setting
Tubing life ≈ 50 – 1000 hours depending on material and speed Norprene longest; silicone shortest and cheapest

Matching Flow to Tubing and Speed

Flow is the product of bore and rotational speed, so pick the target flow first and then choose a tube that reaches it in the comfortable middle of the speed range. A bore that needs full speed to hit your set point will wear out fast and dose unevenly; a very wide bore crawling at minimum speed gives you no fine control. As a starting guide, 1.6 millimetre bore spans roughly 0.5 to 30 millilitres per minute, 3.2 millimetre bore about 2 to 120 millilitres per minute, and 4.8 millimetre bore several hundred millilitres per minute. Viscous liquids lower the achievable flow and raise the pressure the pump must overcome, so step up one bore size and accept a lower speed.

Tubing Materials at a Glance

Material Chemical Range Pumping Life Typical Use
Silicone Aqueous, biological; poor with solvents Short to moderate General media and reagent transfer, autoclavable
Platinum-cured silicone Aqueous, biological Moderate Cell culture and pharmaceutical work, low extractables
Norprene / TPE Broad, including many solvents Longest Continuous duty, perfusion, industrial transfer
Tygon / PVC General reagents and buffers Moderate Clear tubing where visual inspection helps
Viton / fluoropolymer Aggressive solvents, acids, oxidisers Moderate to long Harsh chemistry, at higher cost

Calibration and Dosing Accuracy

A pump is only as accurate as its last calibration. Set the pump to deliver into a tared vessel for a fixed time at your working speed, weigh the delivered liquid, convert using density, and repeat three times. Then adjust the set point by the average ratio. Because output falls steadily as the tube fatigues, calibrate when the tube is new, log the value, and repeat at a fixed interval — weekly for critical dosing, monthly for general transfer. Better still, log the calibration so a shift in delivered volume is the signal to replace tubing rather than a mystery in your results.

Maintenance and Practical Rules

Ordering Checklist

Tell us what you are pumping, the flow you need and how often the pump will run, and we will specify the pump head, tubing grade and spare set that fits the job, with local delivery in Singapore and consumable tubing held in stock.

Contact JWRESEARCH

JW RESEARCH SUPPLY — laboratory equipment and consumables, Singapore.
Telephone / WhatsApp: +65 8039 3660
Email: SINGAPORE@JWSHY.COM
Website: www.jwresearch.cn